Chapter 3 Monitor Calibration
22
English
How to Save the Calibration Setting in a File
Several calibration settings are available after a calibration setting is saved in a file and
saved in an optional folder.
[Procedure]
1. Clic
k [File] - [Save as] after changing calibration settings.
2. Specify a place, name the file and click [OK].
Click [File] - [Open] to open the file when the saved file needs to be read into the
calibration setting window.
Sensor Setting
[Procedure]
1. When clicking [Configuration] - [Change Settings] on the menu bar, the setting
dialog appears. Click “Sensor” and then specify a sensor for calibration.
When a selected sensor is available in communication, “xxx is connectedappears on
the status bar of the main screen (when a selected sensor is not available, “Failed to
communicate with sensor.” message appears).
Tips
A sensor name goes in xxx.
The link state of the current sensor is reconfirmed when it is reselected.
When the Integrated Front Sensor/Swing Sensor is connected, the message is not
displayed on the status bar.
Note
RadiCS software recognizes only the sensor detected first when EIZO sensor are
connected to the computer at the same time (the sensor RadiCS recognizes will be
displayed at the bottom of the main screen). To use another sensor, uninstall all the
sensors and connect only one sensor to the computer.