RadiCS
Chapter 4 Acceptance Test / Constancy Test
39
English
Executing Constancy Test
Pattern Check
Displays test patterns and allows an operator to see if specific pattern classification is
possible with the human eye.
For the monitors with the following conditions, the simultaneous pattern check for
multiple monitors is available.
• Monitors that are made by the same manufacturer and that are the same model.
• EIZO monitors with CAL Switch function that are the same model and have the same CAL
mode.
•
Monitors that have the same standard (including Custom).
Note
• Pattern addition, setup, delete, sequence change, and etc. must be performed in the advanced mode.
[Procedure]
1. Click [Task] - [Pattern Check].
Tips
• In the user mode, the pattern check can also be executed by clicking [Proceed] for the
pattern check as displayed on the main menu.
2. “Input tester name” will appear. Input tester’s name and click [OK].
Note
• Input tester’s name in 31 characters.
3. If any monitors meet the conditions for performing the pattern check for
multiple monitors, the following message will appear; “Do you want to use the
simultaneous pattern check for multiple monitors that are the same model and
have the same standard?”.
Click [Yes] to execute the pattern check for multiple monitors.
4. Click [Proceed].
Tips
• To skip the current monitor pattern check and move onto the next monitor, click [Skip].
5. If more than one pattern set is set, select the pattern set to be used on the
pattern set selection screen. When executing the pattern check for multiple
monitors, the setting of the monitor that has the minimum screen number is
displayed.
Tips
• For GS521-ST, if Pattern Check has not been executed, a black image is displayed in full screen.
Comments to this Manuals