RadiCS
Chapter 10 Information
86
English
Grayscale Check
ACR includes an equality sign but RadiCS doesn’t because the target error rate is < 10% of GSDF.
This is a judgment condition for DICOM Part 14 GSDF. The number of grayscale measuring points
is fixed at 18 and is unchangeable. The measurement result is 17 points because it is expressed as
(JNDn + 1 – JNDn) / 2.
Uniformity Check
ACR includes only a notation about chromaticity uniformity of
Δ
(u’, v’). However, since it is
considered necessary to check luminance uniformity as well, RadiCS adds judgment conditions for
luminance uniformity. These are identical to AAPM in content. See the AAPM item for details of the
correlation with RadiCS.
Sensors
ACR contains nothing in particular about sensors or measuring devices. Since this standard was
compiled using AAPM as a reference, sensors are handled in the same manner as AAPM. See the
AAPM item for details of the correlation with RadiCS.
Multiple Monitors
ACR includes a determination for multiple monitors (note: chromaticity determination only, none
for luminance), but by default RadiCS is set not to make such a determination. If necessary, make
settings as indicated in the table above. ACR includes an equality sign but RadiCS does not.
Cautions
ACR is an educational tool to supply physicians, technicians, and physicists with extensive
knowledge related to digital mammography image quality. It is not an implementation standard, a list
of essential requirements, or a quality control standard. For this reason it does not cover the concepts
of acceptance tests or constancy tests. Nevertheless, as a supporter of the ACR guidelines, we offer
a more practical approach, supplementing the deficiencies related to quality control by referring to
AAPM, which is frequently mentioned in ACR.
IEC, JIS
RadiCS Setup
Constancy Test
Primary
Pattern Check Black
Test Pattern 1
Test Pattern 2
Test Pattern 3
Test Pattern 4
Luminance Check
Δ
L’max < 20% baseline value xxx cd/m
2
Δ
L’max < 25% baseline value xxx cd/m
2
Δ
Lamb < 25% baseline value xxx cd/m
2
Grayscale Check
Uniformity Check