RadiCS
Chapter 10 Information
78
English
IEC 61223-2-5: 1994
“Evaluation and routine testing in medical imaging departments - Part2-5: Constancy tests - Image
display devices” formulated by the International Electrotechnical Commission.
Tips
“IEC” written on RadiCS means “IEC 61223-2-5”.
IPEM Report 91: 2005
“Recommended Standards for the Routine Performance Testing of Diagnostic X-ray Imaging
Systems” formulated by Institute of Physics and Engineering in Medicine in the UK. It applies to
diagnostic X-ray imaging systems as a whole including image display devices but does not include
MR or ultrasonic systems. The items related to monitors were added when this standard was revised
from Report 77. It mainly defines constancy tests.
Tips
“IPEM” written on RadiCS means “IPEM Report 91”.
DIN V 6868-57: 2001
“Image Quality Assurance in X-ray Diagnosis - Part 57: Acceptance testing for image display
devices” formulated by the German Institute for Standardization (Deutsches Institut für Normung
e.V). Image display devices are divided into 3 categories. “Application Category A” includes image
display devices used for the diagnosis of images of high spatial and contrast resolution. “Application
Category B” includes image display devices for diagnosis which are not classified in “Application
Category A” and image display devices for image viewing.
Tips
“DIN” on RadiCS means “DIN V 6868-57”.
Qualitätssicherungs-Richtlinie (QS-RL): 2007
“Guideline for implementing quality assurance of the X-ray systems for diagnostic and medical
treatment purposes according to the chapter 16 and 17 of the X-ray Ordinance”. This defines
the details of the quality assurance of general X-ray systems obliged by the X-ray Ordinance (for
diagnostic: chapter 16, for medical treatment: chapter 17). DIN V 6868-57 is supposed to be referred
on basic test methods for diagnostic image display devices. Limiting values such as the minimum
value of the maximum luminance and the items/frequency of the constancy test are added to the
contents of DIN V 6868-57 that defines only the acceptance test. Although the classification of
image display devices conforms to DIN V 6868-57 (Category A, B), stricter criteria are established
for mammography equipments by reference to PAS1054 “Requirements and testing of digital
mammographic X-ray equipment”, which is the standard issued by the German Institute for
Standardization.
Tips
“QS-RL” written on RadiCS means “Qualitätssicherungs-Richtlinie”. In addition, the indication
“Application Category A Mammo” signifies compliance with PAS1054.